Stat Fax 2100 Automatic Micro Plate Reader
Features
- Fully automatic
- Bichromatic optics
- On board data reduction calculations include regressions, curve fitting and cutoff absorbance calculations with step-by-step user prompting
- Large, non-volatile memory stores at least fifty user entered tests
- Includes parallel and serial outputs (serial cable included)
- Operates with or without a computer. A bidirectional inter-face is available as an accessory for linking to either a Macintosh or IBM type computer
Options & Accessories
- Four and six filter models are available
- Optional 340nm reading capability
- Computer interface for Macintosh or IBM type PC's (2100-A-MAC, 2100-A-PC)
- Serial printers for either 110 or 220V AC
- Dri-Dye® check strips for verifying accuracy, linearity, and repeatability
Technical pecifications
Photometric
| |
Linear measurement range: |
-0.20 to 3.0 absorbance units (A) |
| Photometric accuracy: |
±1% or better |
| Stability: |
drift of no more than 0.005A in 8 hours |
| Light source: |
tungsten lamp with lamp saver feature |
| Wavelengths: |
- Standard
- 405,450,492, and 630nm
- 6-VIS
- 405,450,492,545,600 and 630nm
- 6-UV
- 340,405,450,492,545, and 630nm
|
| Filter type: |
3 cavity sealed interface, 10nm half bandpass |
| Vessel: |
standard 96-well microtiter plates, or strip trays |
Electronic
|
| Display: | two line LCD, alpha numeric, 24 characters per line |
| Printer requirements: | parallel or serial, 80 Column, 2K byte minimum buffer capacity |
| Keyboard: | 29-key, domed membrane switch, enunciating |
| Power requirements: | 115V or 230V AC, 50-60Hz (switch selectable) |
| Serial port: | 2400 baud, 1 start bit, 8 data, 1 stop, no parity |
| Microprocessor: | Z80a or Z180 |
| Non-volatile memory: | battery supported RAM, holds approx. 50 tests and curves |
| UL Listed |
Software
|
| Speed: | reads, and prints absorbances of 96 wells in about two minutes |
| Calculation Modes: | single point calibration, uptake mode, point to point curve fit, polynomial regression linear regression(log & linear), cutoff absorbance |
| Additional features: | complete user prompting, flags and error messages, partial plate reading, alphanumeric test naming, automatic interpretation options, duplicate options, clock and calendar, controls locator, self check system, curve plotting and editing |
Other
|
| Mixer: | variable time, fixed speed |
| Enclosure: | fire-retardant ABS plastic cover with metal base |
| Dimensions: | approx.17x14.5x7in (43x37x18cm) weighs 30 lbs. (13.7 kg.) |
| |