Home
Antibodies
Antigens
EIA/ELISA
Diagnostic Tests
One-Step Card /
 Membrane Tests
Instumentation

TKA 310 Line of Fully Automated Analyzers for EIA/ELISA Assays

TKA 310 Line of Fully Automated Analyzers for EIA/ELISA Assays Laboratories that perform few analyses can afford to equip themselves with one of these “walk-away” analyzers, thus avoiding the use of several machines, helping reduce costs, save room and protect against human error. The programs allow the creation of profiles containing up to 8 assays to be carried out in a single session. It is also possible to carry out different assays on the same microplate, provided that such assays have the same temperature and incubation schedule. The two models in this line—the TKA 312 and TKA 313—, perform similarly, they only differ in the number of operated microplates: two and three respectively. Like all TKA Analyzers the software of these units allow the storage of Standard Curves that can be retrieved later and used to adjust one or two Calibrators, reducing analysis cost. Reagent containers are included in the kit.

Technical Specifications


Configuration

TKA 312

TKA 313
Standard washable needles:internally and externally coated in order to minimize the carry-over
Number of needles:1
Liquid level sensor:independent for each needle
Sample minimum volume:10 µl
pipetting precision and accuracy:< 1% CV at 10% syringe volume, based on gravimetric evaluation
Carry-over of washable needles:10-6 con sufficient needle washing
Conductive disposable tips:option
Pump for needle washing:1
Minimum no. of assays for Profile:8
Maximum number of microplates:23

Washer

TKA 312

TKA 313
Manifold:8 distribution channels and 8 aspiration channels
Number of washing buffers:3
Dispensing speed:2, programmable
Soak time:programmable
Strip position:automatic control of strip absence/presence

Incubators

TKA 312

TKA 313
Number of independent Incubators:23
Type:closed and darkened
Temperature:fixedfrom 2ºC to 50ºC in 0.1ºC steps
Temperature accuracy:± 0.2ºC
Incubation Time:from 1 minute to 24 hours, programmable
Microplate shaking:at intervals, programmable
Constant microplate shaking:option

Reader

TKA 312

TKA 313
Wavelength range:400 - 750 nm (320 - 800 nm on request)
Number of standard filters:4
Standard filters:405, 450, 492, and 620 nm
Max. no. of filter positions:up to 6, automatically selected
Accuracy:< 1% at 450 nm 1.0 0D
Linearity:± 1% at 450 nm
Reproducibility:< ± 0.3% at 450 nm, 1.0 0D (using neutral filter
Range:from 0 to 3 0D
Reading speed:< 5 seconds for single wavelength measurements
Kind of reading:single/dual wavelength measurements

Miscellaneous

TKA 312

TKA 313
Power:110/230 V, 50/60 Hz
Absorption:450 watts500 watts
Working Temperature:15ºC - 35ºC (59ºF - 95ºF)
Relative Humidity:10% - 85% at 30ºC (85ºF) or below
Length:900 mm1070 mm
Depth:700 mm
Height:700 mm
Computer configuration:Hard disk, floppy, CD-ROM, keyboard, color monitor and mouse incl.
Interfaces:RS 232 and parallel
Options:conductive disposable tips, bar code reader