TKA 500 Line of Fully Automated Analyzers for EIA/ELISA Assays

TKA 500 Photo 1

TKA 500 Photo 2

TKA 540 Line

A high performance and profitability line, due to simultaneous action of 4 needles that distribute 96 samples in the microplate within a very short period. This line includes three models—TKA 543 that operates 3 microplates, TKA 544 that operates 4 microplates and TKA 546 that operates 6 microplates. In order to avoid contamination each needle is fitted with its own membrane pump for internal washing purpose. The operator can freely choose up to a maximum of twelve assays to be included in the Profile to carry out. These units are an adequate solution for medium-size laboratories.

TKA 580 Line

TKA 586 with 6 microplates and TKA 589 with 9 microplates are two top models of the new line and are particularly adequate to perform routine screening tests in laboratories having huge workloads and in blood banks, especially due to the use of eight needles. A whole microplate is processed in approximately 5 minutes (100 μl in individual quotas). Each incubator can freely programmed at an individual temperature. Futhermore, TKA 589 can be provided with the “STAT function" option that allows samples and new analysis, if so required, during the session.

Technical Specifications


Configuration

TKA 543 - 544

TKA 546

TKA 586

TKA 589
Standard washable needles:internally and externally coated in order to minimize the carry-over
Number of needles:48
Liquid level sensor:independent for each needle
Sample minimum volume:10 µl
pipetting precision and accuracy:< 1% CV at 10% syringe volume, based on gravimetric evaluation
Carry-over of washable needles:10-6 con sufficient needle washing
Conductive disposable tips:option
Pump for needle washing:48
Minimum no. of assays for Profile:12
Maximum number of microplates:3 - 469

Washer

TKA 543 - 544

TKA 546

TKA 586

TKA 589
Manifold:8 distribution channels and 8 aspiration channels
Number of washing buffers:4
Dispensing speed:2, programmable
Soak time:programmable
Strip position:automatic control of strip absence/presence

Incubators

TKA 543 - 544

TKA 546

TKA 586

TKA 589
Number of independent Incubators:3 - 469
Type:closed and darkened
Temperature:from 2ºC to 50ºC in 0.1ºC steps
Temperature accuracy:± 0.2ºC
Incubation Time:from 1 minute to 24 hours, programmable
Microplate shaking:at intervals, programmable
Constant microplate shaking:option

Reader

TKA 543 - 544

TKA 546

TKA 586

TKA 589
Wavelength range:400 - 750 nm (320 - 800 nm on request)
Number of standard filters:5
Standard filters:405, 450, 492, 620 and 690 nm
Max. no. of filter positions:up to 6, automatically selected
Accuracy:< 1% at 450 nm 1.0 0D
Linearity:± 1% at 450 nm
Reproducibility:< ± 0.3% at 450 nm, 1.0 0D (using neutral filter
Range:from 0 to 3 0D
Reading speed:< 5 seconds for single wavelength measurements
Kind of reading:single/dual wavelength measurements

Miscellaneous

TKA 543 - 544

TKA 546

TKA 586

TKA 589
Power:110/230 V, 50/60 Hz
Absorption:600 watts650 watts750 watts850 watts
Working Temperature:15ºC - 35ºC (59ºF - 95ºF)
Relative Humidity:10% - 85% at 30ºC (85ºF) or below
Length:1350 mm1670 mm2070 mm
Depth:790 mm
Height:810 mm
Computer configuration:Hard disk, floppy, CD-ROM, keyboard, color monitor and mouse incl.
Interfaces:RS 232 and parallel
Options:conductive disposable tips, bar code reader