International
Immuno-Diagnostics
This Chemistry Analyzer comes equipped with brilliant bichromatic optics, on-board curve-fitting software, and an integrated printer to produce high throughput and enhance productivity. On-board curve fitting calculations are accompanied by step-by-step user prompting to simplify and make operation easier. Including an interactive touch-screen LCD with a USB mouse option, it also has a large non-volatile memory that stores up to 120 user-entered tests. This chemistry analyzer even performs single point calibration by standard or factor, multipoint calibration and kinetic rate by standard or factor (batch or single). Operating with IAD filters, the Stat Fax 4500 accommodates 12 mm tubes and cuvettes. The economical unit also has a read cell which is temperature controlled at 37 °C. Standard models include six filters (340 nm, 405 nm, 505 nm, 545 nm, 580 nm, and 630 nm).
Stat Fax® 4500 Chemistry Analyzer
Options & Accessories
Linear Measurement Range:
Tube or Cuvette: 0.0 to 3.0 Absorbance units (A)
Photometric Accuracy:
+/-(1% of the reading + 0.005A
Stability:
Drift of no more than 0.005A in 8 hours/bichromatic
Light Source:
Tungsten lamp with lamp saver feature
Standard Wavelengths:
340, 405, 505, 545, 580, and 630nm (alternate filters available from 340 to 700nm)
Filter Type
IAD hardcoat interference, 10nm half bandpass
Tube Size:
12mm round is standard
Minimum Sample Volume:
1mL for 12mm round tube
Technical Specifications
Software
Interactive touch-screen LCD with USB mouse option
Bichromatic optics
On board curve fitting calculations and various calibration points
Flowcell to accomodate 12mm tubes or cuvettes
IAD filters
Photometric
Electronic
Other
Display:
Interactive touch-screen 3.5" LCD, color graphic display
Printer:
Thermal dot matrix, with graphic capability
Power Requirements:
115V or 230V AC, 1.5A, 50-60Hz (universal input)
Microprocessor
eZ80
Interface:
USB mouse
Speed:
Reads, calculates and prints results, 3 seconds per tube
Calculation Modes:
Single point calibration by standard or factor, multipoint calibration with point-to-point curve fit, rate by standard or factor (batch or singly)
Test Menu:
More than 50 open channels to store tests
Stores all parameters including wavelengths, calculations, unit codes. linear and normal ranges, rate timing, standrd values, test names and previous standard curve
Complete user prompting, flags and error messages, partial plate reading, alphanumeric test naming, automatic interpretation options, duplicate options, clock and calendar, controls locator, self-check system, curve plotting and editing
Temperature Control:
Read cell and incubation block, user selectable
Enclosure:
Painted flame-retardant ABS plastic cover and base
Dimensions:
Approx. 9 x 13 x 5in (24 x 34 x 13cm), weighs 10 lbs. (4.5 kg)
Certifications:
NRTL listed, CE mark