This Chemistry Analyzer comes equipped with brilliant bichromatic optics, on-board curve-fitting software, and an integrated printer to produce high throughput and enhance productivity. On-board curve fitting calculations are accompanied by step-by-step user prompting to simplify and make operation easier. Including an interactive touch-screen LCD with a USB mouse option, it also has a large non-volatile memory that stores up to 120 user-entered tests. This chemistry analyzer even performs single point calibration by standard or factor, multipoint calibration and kinetic rate by standard or factor (batch or single). Operating with IAD filters, the Stat Fax 4500 accommodates 12 mm tubes and cuvettes. The economical unit also has a read cell which is temperature controlled at 37 °C. Standard models include six filters (340 nm, 405 nm, 505 nm, 545 nm, 580 nm, and 630 nm).

Stat Fax® 4500 Chemistry Analyzer

Options & Accessories

Linear Measurement Range:

Tube or Cuvette: 0.0 to 3.0 Absorbance units (A)

Photometric Accuracy:

+/-(1% of the reading + 0.005A

Stability:

Drift of no more than 0.005A in 8 hours/bichromatic

Light Source:

Tungsten lamp with lamp saver feature

Standard Wavelengths:

340, 405, 505, 545, 580, and 630nm (alternate filters available from 340 to 700nm)

Filter Type

IAD hardcoat interference, 10nm half bandpass

Tube Size:

12mm round is standard

Minimum Sample Volume:

1mL for 12mm round tube

Please reload

Technical Specifications
Software

Interactive touch-screen LCD with USB mouse option

Bichromatic optics

On board curve fitting calculations and various calibration points

Flowcell to accomodate 12mm tubes or cuvettes

IAD filters

Please reload

Photometric
Electronic
Other

Display:

Interactive touch-screen 3.5" LCD, color graphic display

Printer:

Thermal dot matrix, with graphic capability

Power Requirements:

115V or 230V AC, 1.5A, 50-60Hz (universal input)

Microprocessor

eZ80

Interface:

USB mouse

Please reload

Speed:

Reads, calculates and prints results, 3 seconds per tube

Calculation Modes:

Single point calibration by standard or factor, multipoint calibration with point-to-point curve fit, rate by standard or factor (batch or singly)

Test Menu:

More than 50 open channels to store tests

Stores all parameters including wavelengths, calculations, unit codes. linear and normal ranges, rate timing, standrd values, test names and previous standard curve

Complete user prompting, flags and error messages, partial plate reading, alphanumeric test naming, automatic interpretation options, duplicate options, clock and calendar, controls locator, self-check system, curve plotting and editing

Please reload

Temperature Control:

Read cell and incubation block, user selectable

Enclosure:

Painted flame-retardant ABS plastic cover and base

Dimensions:

Approx. 9 x 13 x 5in (24 x 34 x 13cm), weighs 10 lbs. (4.5 kg)

Certifications:

NRTL listed, CE mark