This instrumnet is designed to read and calculate the results of in-vitro diagnostic assays. Equipped with a large graphical display, the reader provides absorbance and concentration readings at or near the available wavelength. It features multiple, self-prompting operating modes to meet the varying demands and requirements of any facility, and accepts standard 12 mm round tubes as well as 1 cm square cuvettes. Basic calculations are also permanently stored in the Stat Fax memory, including several single and multi-point equations.

Stat Fax® 3300 Chemistry Analyzer

Options & Accessories

Linear Measurement Range:

0.0 to 3.0 Absorbance units (A)

Photometric Accuracy:

+/-(1% +/- 0.010) 0.0 through 1.5 Abs

+/-(2% +/- 0.010) 1.5 through 3.0 Abs

Stability:

Drift of no more than 0.005A in 8 hours/bichromatic

Light Source:

Tungsten lamp with lamp saver feature

Standard Wavelengths:

Standard: 405, 450, 492, and 630nm

6-VIS: 405, 450, 492, 545, and 630nm

6-UV: 340, 405, 450, 492, 545, and 630nm (alternate filters available from 240 to 700nm)

Filter Type

IAD hardcoat interference, 10nm half bandpass

Vessel:

Non-break-apart or break-apart strips up to 12 wells long, 3-strip loading capacity

Transport Mechanism

Automatic strip transport

Please reload

Technical Specifications
Software

Large graphical display

Built-in printer and connectivity to external printer/ PC

Adapts to use 12 mm round tubes- 1 cm square cuvettes- or built in flow cell

Conforms to GLP/GMP and ISO 9001 Standard

Please reload

Photometric
Electronic
Other

Display:

Interactive touch-screen 3.5" LCD, color graphic display

Printer:

Thermal on-board printer with graphic capability

Power Requirements:

100V - 240V AC, 1.2A, 50-60Hz (universal input)

Interface:

USB mouse

Please reload

Speed:

Reads, calculates and prints results, 30 seconds per 12-well strip

Calculation Modes:

Single point calibration by standard or factor, multi-point calibration with point-to-point curve fit, -linear regressions with log, lin, and log-logit selections

Test Menu:

120 open channels to store tests

Stores all parameters including wavelengths, calculations, unit codes, ranges, interpretations, calibrator values, test names, and previous curve

Please reload

Enclosure:

Flame-retardant ABS plastic cover and base

Dimensions:

Approx. 9 x 13 x 5.5in (24 x 34 x 13cm), weighs 5 lbs. (4.5 kg)

Certifications:

NRTL listed, CE mark